Clinical performance evaluation of Streptococcus pneumoniae antigen detection kit ODK0501 in childhood respiratory tract infections
Satoshi IWATA1,2), Keita MATSUBARA1), Naohisa KAWAMURA3), Koichi SHIMIZU3), Kohji AZUMAGAWA4), Takao MORINOBU5), Shinobu ISHIZAWA6), Reiko TAKAYANAGI6), Mika NUMATA6), Hiroshi UCHIDA7), Kumiko SUGITA8), Toshio TAKASHIMA9), Chikara NAKAHAMA10), Suguru AKAMATSU11), Keisuke SUNAKAWA12)
A diagnostic kit (ODK0501) for detecting a Streptococcus pneumoniae (S. pneumoniae) antigen was evaluated using rhinopharyngeal swabs from 105 children with pneumonia or acute bronchitis. ODK0501 has been designed to specifically detect C-polysaccharide (C-ps), a cell wall antigen of S. pneumoniae. Sensitivity, specificity, and concordance rate were 66.0% (33/50), 100% (55/55), and 83.8% (88/105), respectively, demonstrating the usefulness of the kit as a rapid assay for pneumococcal respiratory tract infections in children.
1) Department of Pediatrics, National Hospital Organization Tokyo Medical Center
2) Center for Infectious Diseases and Infection Control, Keio University School of Medicine
3) Department of Pediatrics, Osaka Rosai Hospital
4) Department of Pediatrics, Seikeikai Hospital
5) Department of Pediatrics, Saiseikai Ibaraki Hospital
6) Department of Pediatrics, Tohoku Rosai Hospital
7) Uchida Children's Clinic
8) Sugita Children Clinic
9) Takashima Shonika Iin
Key words | |
---|---|
Received | July 23, 2010 |
Accepted | September 9, 2010 |
23 (1):3─9,2011
- Vol. 36
- Vol. 35
- Vol. 34
- Vol. 33
- Vol. 32
- Vol. 31
- Vol. 30
- Vol. 29
- Vol. 28
- Vol. 27
- Vol. 26
- Vol. 25
- Vol. 24
- Vol. 23
- Vol. 22
- Vol. 21
- Vol. 20
- Vol. 19
- Vol. 18